Remote Control Commands
R&S
®
RTB2000
335User Manual 1333.1611.02 ─ 04
Parameters:
<Mode> YT | XY
YT
Default time diagram with a time axis in x-direction and the sig-
nal amplitudes displayed in y-direction.
XY
XY-diagram, combines the voltage levels of two waveforms in
one diagram.
*RST: YT
DISPlay:XY:XSOurce <Source>
Defines the source to be displayed in x-direction in an XY-diagram, replacing the usual
time base.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel R&S RTB2000
oscilloscopes.
*RST: CH1
DISPlay:XY:Y1Source <Source>
Defines the first source to be displayed in y-direction in an XY-diagram.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel R&S RTB2000
oscilloscopes.
*RST: CH2
DISPlay:XY:Y2Source <Source>
Defines an optional second source to be displayed in y-direction in an XY-diagram.
The command is only relevant for 4-channel R&S RTB2000 instruments.
Parameters:
<Source> NONE | CH1 | CH2 | CH3 | CH4
*RST: NONE
15.8.4 Digital Voltmeter
The DVM suffix <m> sets the number of the DVM measurement (measurement place).
Applications