Logic Analyzer (Option R&S RTB-B1, MSO)
R&S
®
RTB2000
221User Manual 1333.1611.02 ─ 04
4 = scales all visible channels to a maximum
5 = shows the threshold of the channels
6 = shows the activity of the logic channel
Logic channels - activity display
The activity symbols of the logic channel show the current status of all logic channels
and can have the following values:
: logic channel is low
: logic channel is high
: a change in the state of the logic channel has occurred during the measuring
interval
Remote commands:
DIGital<m>:CURRent:STATe:MINimum? on page 430
DIGital<m>:CURRent:STATe:MAXimum? on page 430
LOGic<p>:CURRent:STATe:MINimum? on page 430
LOGic<p>:CURRent:STATe:MAXimum? on page 430
13.2 Logic Analyzer Settings
To activate the logic analyzer
Prerequisite: a logic probe is connected to the instrument.
1. Press the LOGIC key.
The "Logic" menu opens.
2. Press "Logic Probe" to select the logic channels group you want to activate.
3. Enable the "State".
Logic Analyzer Settings