EasyManua.ls Logo

Rohde & Schwarz R&S RTB2004 - Width Search

Rohde & Schwarz R&S RTB2004
509 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waveform Analysis
R&S
®
RTB2000
90User Manual 1333.1611.02 ─ 04
6.5.4 Width Search
The width search finds pulses with an exact pulse width, or pulses shorter or longer
than a given time, or pulses inside or outside the allowable time range. It is similar to
the width trigger.
SEARCH > "Search Type" = "Width" > "Setup"
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
SEARch:TRIGger:WIDTh:POLarity on page 285
Level
Sets the voltage level on which the pulse width is measured. To let the instrument set
the level, tap "Find Threshold".
Remote command:
SEARch:TRIGger:WIDTh:LEVel on page 285
Hysteresis
Sets a hysteresis range to the search level to avoid unwanted search results caused
by noise oscillation around the level. To let the instrument set the hysteresis, tap "Find
Threshold".
Remote command:
SEARch:TRIGger:WIDTh:LEVel:DELTa on page 285
Comparison
Sets the condition how the measured pulse width is compared with the given limits.
Search