Remote Control Commands
R&S
®
RTB2000
417User Manual 1333.1611.02 ─ 04
Parameters:
<DataLength> Range: 0 to 8
Increment: 1
*RST: 1
Default unit: Byte
SEARch:PROTocol:CAN:DCONdition <DataCondition>
Sets the comparison condition for data: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
The command is relevant if SEARch:PROTocol:CAN:CONDition is set to IDData.
Parameters:
<DataCondition> EQUal | NEQual | GTHan | LTHan
*RST: EQU
SEARch:PROTocol:CAN:DATA <Data>
Defines the data pattern. The pattern length is defined with SEARch:PROTocol:CAN:
DLENgth.
The command is relevant if SEARch:PROTocol:CAN:CONDition is set to IDData.
Parameters:
<Data> String containing binary pattern with max. 64 bit. Characters 0,
1, and X are allowed. Make sure to enter complete bytes.
15.11.6 LIN (Option R&S RTB-K3)
The Local Interconnect Network (LIN) is a simple, low-cost bus system used within
automotive network architectures.
Note: SPI/SSPI and UART protocols occupy two bus lines (bus 1 and 2 or bus 3 and
4). If one of these buses is configured, the number of buses (suffix <b>) is reduced.
Bus 2 and/or bus 4 is not available.
LIN - Configuration................................................................................................417
LIN - Trigger..........................................................................................................419
LIN - Decode Results............................................................................................421
LIN - Search..........................................................................................................427
15.11.6.1 LIN - Configuration
Start the bus configuration with the threshold setting. Use one of the following com-
mands:
CHANnel<m>:THReshold:FINDlevel on page 256
Serial Bus Analysis