Remote Control Commands
R&S
®
RTB2000
318User Manual 1333.1611.02 ─ 04
Parameters:
<State> ON | OFF
*RST: OFF
CURSor<m>:SOURce <Source>
Defines the source of the cursor measurement.
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4 | MA1 | RE1 | RE2 | RE3 | RE4 | XY1 |
XY2 | D70 | D158 | D0..15
CH1 | CH2 | CH3 | CH4
Active analog channel waveform 1 to 4
MA1
Active math waveform
RE1 | RE2 | RE3 | RE4
Active reference waveform 1 to 4
XY1 | XY2
Active XY-waveform
D0..15
Active digital channels D0 to D15, available if MSO option
R&S RTB-B1 is installed.
On individual digital channels, only vertical (time) cursor mea-
surements are possible.
D70 | D158
Active digital channels D0...D7 (pod 1) and D8...D15 (pod 2).
Only available, if MSO option R&S RTB-B1 is installed. On pods,
only V-marker measurement is possible.
CURSor<m>:FUNCtion <Type>
Defines the cursor measurement type.
Suffix:
<m>
.
1, the suffix is irrelevant.
Parameters:
<Type> HORizontal | VERTical | HVERtical
To set the V-marker measurement, use CURSor<m>:
TRACking[:STATe].
*RST: VERTical
Measurements

Table of Contents