Remote Control Commands
R&S
®
RTB2000
395User Manual 1333.1611.02 ─ 04
Parameters:
<IdleState> IDLLow | IDLHigh
IDLLow: idle low, low = 1
IDLHigh: idle high, high = 1
*RST: IDLH
BUS<b>:UART:DATA:POLarity <Polarity>
Defines if the transmitted data on the bus is high (high = 1) or low (low = 1) active.
Alternative command for BUS<b>:UART:POLarity .
Suffix:
<b>
.
1 | 2
Parameters:
<Polarity> POSitive | NEGative
POSitive = high active
NEGative = low active
*RST: POS
BUS<b>:UART:SSIZe <SymbolSize>
Sets the number of data bits in a message.
Suffix:
<b>
.
1 | 2
Parameters:
<SymbolSize> Range: 5 to 9
Increment: 1
*RST: 8
Default unit: Bit
BUS<b>:UART:PARity <Parity>
Defines the optional parity bit that is used for error detection.
Suffix:
<b>
.
1..2
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Parity> ODD | EVEN | NONE
*RST: NONE
BUS<b>:UART:SBIT <StopBitNumber>
Sets the stop bits.
Serial Bus Analysis

Table of Contents