Trigger
R&S
®
RTB2000
60User Manual 1333.1611.02 ─ 04
Figure 5-2: Pulse width is shorter (left) or longer (right) than a given duration (also known as glitch
trigger)
min
max
min
max
1 2
Figure 5-3: Pulse width is inside or outside an allowable time range
1 = Inside: min width < pulse < max width
2 = Outside: pulse < min width OR pulse > max width
±Δ
1 2
±Δ
Figure 5-4: Pulse width is equal or unequal to a given duration, with optional variation (
Δ
)
1 = Equal: (width - variation) < pulse < (width + variation)
2 = Unequal: pulse < (width - variation) OR pulse > (width + variation)
TRIGGER > "Trigger Type" = "Width"
Width Trigger