Measurements
R&S
®
RTM20xx
86User Manual 1317.4726.02 ─ 01
The following measurement results are determined for the displayed section of the wave-
form:
Label Description Display
Vp+ Positive peak value WF
Vp- Negative peak value WF
tr Rising time of the first rising edge WF
Mean Mean value WF
tf Falling time of the first falling edge WF
RMS RMS T
Vpp Peak to peak value T
T Period length T
f Frequency T
When you activate quick measurements, cursor measurements are automatically deac-
tivated, as well as the reference and math menus. Deactivate quick measurements before
selecting these functions. Channels other than the selected one are switched off in quick
measurement mode.
► Press the QUICK MEAS key again to deactivate the quick measurement and remove
the results.
Quick Measurements