146User Manual 1317.4726.02 ─ 01
Comparison
Sets the condition how the duration of a steady pattern is compared with the given limit(s).
The three settings "Width" "Variation" and "Comparison" define the time range for keeping
up the true result of the state pattern.
The same conditions as with width search are used, see "Comparison" on page 138.
Remote command:
SEARch:TRIGger:PATTern:WIDTh:RANGe on page 371
Width
Sets the limit time of a steady pattern, the nominal value for comparisons.
Remote command:
SEARch:TRIGger:PATTern:WIDTh[:WIDTh] on page 372
Variation
Sets a range Δt to the reference "Width" if comparison is set to "Equal" or "Not equal".
The instrument finds true results of the state pattern inside or outside the range width ±
Δt.
Remote command:
SEARch:TRIGger:PATTern:WIDTh:DELTa on page 372
10.3.9 Gate Menu
Access: SEARCH MENU > "Gate"
The gate restricts the time base of the source waveform to be searched.
Gate
Defines the search area. If the search is performed on a running acquisition series, the
instrument analyzes the displayed data. The search on a stopped acquisition analyzes
the contents of the memory.
"All"
Running acquisition: all waveform samples that are displayed on the
screen.
Stopped acquisition: all data samples that are stored in the memory.
"Display"
Search is restricted to the time range of the display.
"User"
Search is restricted to the time range defined by "Start" and "Stop" val-
ues.
Remote command:
SEARch:GATE:MODE on page 361
Reference for Search Menu