To set a range Δt, use SEARch:TRIGger:PATTern:WIDTh:DELTa on page 372
Parameters:
<Range> WITHin | OUTSide | LONGer | SHORter
WITHin
Finds patterns steady for a time range width ± Δt.
OUTSide
Finds patterns outside a time range width ± Δt.
LONGer
Finds patterns steady for at least the given width.
SHORter
Finds patterns shorter than the given width.
*RST: LONG
SEARch:
TRIGger:PATTern:WIDTh[:WIDTh] <Width>
Sets the reference time of a steady pattern, the nominal value for comparisons.
Parameters:
<Width> Default unit: s
SEARch:TRIGger:PATTern:WIDTh:DELTa <DeltaTime>
Sets a range Δt to the reference pattern duration set with SEARch:TRIGger:
PATTern:WIDTh[:WIDTh] if SEARch:TRIGger:PATTern:WIDTh:RANGe is set to
WITHin or OUTSide.
Parameters:
<DeltaTime> Default unit: s
15.2.12.9 Search Results
SEARch:RESDiagram:SHOW.........................................................................................372
SEARch:RESult:ALL?.....................................................................................................373
SEARch:RESult<n>?......................................................................................................373
SEARch:RCOunt?..........................................................................................................374
SEARch:RESDiagram:SHOW <ResultShow>
Shows or hides the table of search results.
Parameters:
<ResultShow> ON | OFF
*RST: OFF