Protocol Analysis
R&S
®
RTM20xx
154User Manual 1317.4726.02 ─ 01
11.1.5 Frame Table: Decode Results
Access: PROTOCOL > "Frame Table"
The frame table shows the detailed decoded data for each frame of the acquisition. The
table content is protocol-specific, and the display of the table can be enabled for each
individual bus type.
Decode results shown in the frame table can be saved in a CSV file.
To navigate in the frame table
1. Stop the acquisition.
2. Turn the NAVIGATION knob to select a frame.
3. Press the NAVIGATION knob to display the selected frame in the center of the wave-
form display.
If a search on decoded data was performed, the search results are marked in the frame
table.
Frame Table
Displays or hides the table of decode results.
Remote command:
BUS<b>:RESult on page 376
Track frame
Defines the automatic synchronization of the selected frame in the frame table and the
waveform display.
The function is only available if the acquisition has been stopped.
"Off"
Frame table and waveform display are not synchronized.
"Frame Index"
The waveform display is connected to the frame that is selected in the
frame table. The selected frame is shown in the center of the display.
If you select another frame, the waveform display is adjusted automat-
ically.
"Hori. Position"
The frame selection in the frame table is connected to the waveform
display. The frame in the center of the display is selected in the frame
table. When you change the horizontal position of the waveform, the
selection in the frame table is adjusted automatically.
Frame time difference
If selected, the time shown in the frame table is the time difference to the previous frame.
The column is indicated with "Time diff.". If the setting is disabled, the absolute time in
releation to the trigger point is shown in the "Start time" column.