142User Manual 1317.4726.02 ─ 01
Variation
Sets a range Δt to the reference "Width" if comparison is set to "Equal" or "Not equal".
The instrument finds pulses inside or outside the range width ± Δt.
Remote command:
SEARch:TRIGger:RUNT:DELTa on page 368
10.3.7 Data2Clock Search
The settings for Data2Clock search are provided in two menus: in the "Setup" menu you
define the clock polarity, setup and hold times; and in the "Source Setup" menu you define
the waveforms to be used, and the levels and hysteresis for each source.
10.3.7.1 Data2Clock Source Setup
Access: SEARCH MENU > "Search type = Data2Clock" > "Source Setup"
For Data2Clock search, two sources have to be defined: a clock and a data waveform.
For each waveform, the level and the hysteresis can be set individually.
Clock
Selects the input channel of the clock signal.
Remote command:
SEARch:TRIGger:DATatoclock:CSOurce on page 368
Data
Selects the input channel of the data signal.
Remote command:
SEARch:SOURce on page 361
Level
Set the voltage levels for clock and data signals. Clock level and clock edge define the
reference point for setup and hold time. The data lavel defines the point of data transition.
Remote command:
SEARch:TRIGger:DATatoclock:CLEVel on page 369
SEARch:TRIGger:DATatoclock:DLEVel on page 369
Reference for Search Menu