EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Runt Setup

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Search
R&S
®
RTM20xx
141User Manual 1317.4726.02 ─ 01
Variation
Sets a range Δt to the reference "Rise/Fall time" if comparison is set to "Equal" or "Not
equal". The instrument finds rise/fall times inside or outside the range width ± Δt.
Remote command:
SEARch:TRIGger:RISetime:DELTa on page 366
10.3.6 Runt Setup
Access: SEARCH MENU > "Search type = Runt" > "Setup"
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
SEARch:TRIGger:RUNT:POLarity on page 367
Upper level
Sets the upper voltage threshold for runt detection. A negative runt crosses the upper
level twice without crossing the lower level.
Remote command:
SEARch:TRIGger:LEVel:RUNT:UPPer on page 367
Lower level
Sets the lower voltage threshold for runt detection. A positive runt crosses the lower level
twice without crossing the upper level.
Remote command:
SEARch:TRIGger:LEVel:RUNT:LOWer on page 367
Comparison
Sets the condition how the measured runt width is compared with the given limit(s).
The same conditions as with width search are used, see "Comparison" on page 138.
Remote command:
SEARch:TRIGger:RUNT:RANGe on page 367
Width
Sets the reference runt pulse width, the nominal value for comparisons.
Remote command:
SEARch:TRIGger:RUNT:WIDTh on page 368
Reference for Search Menu

Table of Contents

Related product manuals