SYSTem:ERRor:[NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response consists of an error number and a short description of the error.
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> Error/event_number,"Error/event_description>[;Device-depend-
ent info]"
If the queue is empty, the response is 0,"No error"
Usage: Query only
SCPI confirmed
SYSTem:
ERRor:ALL?
Queries the error/event queue for all unread items and removes them from the queue.
The response is a comma separated list of error number and a short description of the
error in FIFO order.
Positive error numbers are instrument-dependent. Negative error numbers are reserved
by the SCPI standard.
Return values:
<Error> List of: Error/event_number,"Error/event_description>[;Device-
dependent info]"
If the queue is empty, the response is 0,"No error"
Usage: Query only
SCPI confirmed
SYST:PRESet
Resets the instrument to the default state, has the same effect as *RST.
Usage:
Event
15.2.17 Status Reporting
15.2.17.1 STATus:OPERation Register
The commands of the STATus:OPERation subsystem control the status reporting
structures of the STATus:OPERation register:
See also:
chapter 15.1.6.1, "Structure of a SCPI Status Register", on page 262
"STATus:OPERation Register" on page 267