267User Manual 1317.4726.02 ─ 01
Bit No. Meaning
6 User Request
This bit is set when the instrument is switched over to manual control.
7 Power On (supply voltage on)
This bit is set on switching on the instrument.
STATus:OPERation Register
In the CONDition part, this register contains information on which actions the instrument
is being executing. In the EVENt part, it contains information on which actions the instru-
ment has executed since the last reading. It can be read using the commands
STATus:OPERation:CONDition? or STATus:OPERation[:EVENt]?.
See also: figure 15-2
The remote commands for the STATus:OPERation register are described in chap-
ter 15.2.17.1, "STATus:OPERation Register", on page 453.
Table 15-5: Bits in the STATus:OPERation register
Bit No. Meaning
0
ALIGnment
This bit is set as long as the instrument is performing a self alignment.
1
SELFtest
This bit is set while the selftest is running.
2
AUToset
This bit is set while the instrument is performing an auto setup.
3
WTRIgger
This bit is set while the instrument is waiting for the trigger.
4 to 14 Not used
15 This bit is always 0.
STATus:QUEStionable Register
This register contains information about indefinite states which may occur if the unit is
operated without meeting the specifications. It can be read using the commands
STATus:QUEStionable:CONDition? on page 456 and STATus:
QUEStionable[:EVENt]? on page 457