Protocol Analysis
R&S
®
RTM20xx
183User Manual 1317.4726.02 ─ 01
Select Symbol ← Pattern
Selects the symbol for binary or hex input.
Value ← Pattern
Sets the hexadecimal value for the selected symbol by turning the navigation knob.
Parity Error
Triggers on a parity error indicating a transmission error.
Remote command:
TRIGger:A:UART:MODE on page 402 (PERRor)
Frame Error
Triggers on a frame error.
Remote command:
TRIGger:A:UART:MODE on page 402 (FERRor)
Break
Triggers if a start bit is not followed by a stop bit, the data line remains at logic 0 for longer
than a UART word.
Remote command:
TRIGger:A:UART:MODE on page 402 (BREak)
11.4.4 UART/RS-232 Decode Results
You can enable the decoding results in the "Protocol" main menu. "Decode" shows the
decoded values below the waveforms in the format selected with "Display". Additionally,
you can display the binary signal with "Bits".
See also: chapter 11.1.2, "Protocol - Common Settings", on page 149
Additionally, you can display and save a "Frame Table" containing decoded data: frame
number, start time of the frame, data and state of the frame.
See also: chapter 11.1.5, "Frame Table: Decode Results", on page 154
UART/RS-232 Interface (Option R&S RTM-K2)