EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Acquisition and Setup

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
281User Manual 1317.4726.02 ─ 01
Example:
See chapter 15.3.1.2, "Reading Waveform Data in Unsigned Inte-
ger Format", on page 458
Usage: Query only
CHANnel<m>:DATA:YRESolution?
CHANnel<m>:DATA:ENVelope:YRESolution?
CALCulate:MATH<m>:DATA:YRESolution?
REFCurve<m>:DATA:YRESolution?
Return the vertical bit resolution of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined
(FORM UINT, 8|16|32).
Return values:
<Yresolution> For default waveforms, the resolution is 8 bit.
If high resolution, average or filter are set for the waveform, the
resolution is 16 bit.
Example:
See chapter 15.3.1.2, "Reading Waveform Data in Unsigned Inte-
ger Format", on page 458
Usage: Query only
15.2.4 Acquisition and Setup
Starting and Stopping Acquisition.........................................................................281
Time Base.............................................................................................................282
Acquisition.............................................................................................................284
Vertical..................................................................................................................288
Waveform Data.....................................................................................................293
Probes...................................................................................................................296
15.2.4.1 Starting and Stopping Acquisition
RUN.............................................................................................................................281
RUNContinous...............................................................................................................282
SINGle..........................................................................................................................282
RUNSingle....................................................................................................................282
ACQuire:NSINgle:COUNt................................................................................................282
STOP...........................................................................................................................282
RUN
Starts the continuous acquisition.
Usage:
Event