Remote Control
R&S
®
RTM20xx
296User Manual 1317.4726.02 ─ 01
Setting parameters:
<Points> DEFault | MAXimum | DMAXimum
Sets the range for data queries.
DEFault
Waveform samples that are visible as waveform points on the
screen.
MAXimum
All waveform samples that are stored in the memory. Only avail-
able if acquisition is stopped.
DMAXimum
Display maximum: Waveform samples stored in the current wave-
form record but only for the displayed time range. At maximum
waveform rate, the instrument stores more samples than visible
on the screen, and DMAX returns more values than DEF. Only
available if acquisition is stopped.
*RST:
DEFault
Return values:
<Points> Number of data points in the selected range.
Default unit: Samples
Example:
CHAN:DATA:POIN DEF
CHAN:DATA:POIN?;:CHAN2:DATA:POIN?
Returned values: 10416;10416
CHAN:DATA:POIN DMAX
CHAN:DATA:POIN?;:CHAN2:DATA:POIN?
Returned values: 124992;124992
CHAN:DATA:POIN MAX
CHAN:DATA:POIN?;:CHAN2:DATA:POIN?
Returned values: 4194302;4194302
Example:
See chapter 15.3.1.1, "Reading Waveform Data from Memory",
on page 458
15.2.4.6 Probes
PROBe<m>:SETup:ATTenuation[:AUTO]?.......................................................................297
PROBe<m>:SETup:ATTenuation:UNIT............................................................................297
PROBe<m>:SETup:ATTenuation:MANual........................................................................297
PROBe<m>:SETup:BANDwidth?.....................................................................................297
PROBe<m>:SETup:CAPacitance?...................................................................................298
PROBe<m>:SETup:DCOFfset?.......................................................................................298
PROBe<m>:SETup:IMPedance?.....................................................................................298
PROBe<m>:SETup:MODE ............................................................................................299
PROBe<m>:SETup:NAME?............................................................................................299
PROBe<m>:SETup:OFFSwitch.......................................................................................299
PROBe<m>:SETup:TYPE?.............................................................................................299
PROBe<m>:SETup:UOFFset..........................................................................................300
PROBe<m>:ID:BUILd?...................................................................................................300
Command Reference