PROBe<m>:ID:PARTnumber?........................................................................................300
PROBe<m>:ID:PRDate?.................................................................................................301
PROBe<m>:ID:SRNumber?............................................................................................301
PROBe<m>:ID:SWVersion?............................................................................................301
PROBe<m>:SETup:ATTenuation[:AUTO]?
Returns the attenuation of an automatically detected probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<ProbeAttenuation> Range: 1 to 1000
Usage: Query only
PROBe<m>:
SETup:ATTenuation:UNIT <Unit>
Selects the unit that the probe can measure.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Unit> V | A
Firmware/Software: FW 03.700
PROBe<m>:SETup:ATTenuation:MANual <ManualAttenuation>
Sets the attenuation or gain of the probe if the probe was not detected by the instrument.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ManualAttenuation> Range: 1 to 10000
*RST: 1
Firmware/Software: FW 03.700
PROBe<m>:SETup:BANDwidth?
Queries the bandwidth of the probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.