Protocol Analysis
R&S
®
RTM20xx
202User Manual 1317.4726.02 ─ 01
Identifier Setup ← Identifier and data
Opens the "Identifier" menu.
The identifier conditions are the same as for the "Identifier" trigger type, see "Identifier"
on page 200.
Data Setup ← Identifier and data
Opens a menu to set the data pattern. After setting the "No. of Bytes" and the "Com-
pare" condition, you can enter the value bit-by-bit by setting the state high, low, or don't
care for each single bit. Alternatively, you can enter a hexadecimal value for each byte.
The method is the same as for identifier input. See:
"Compare" on page 201
"Bit" on page 201
"State" on page 201
"Byte" on page 201
"Value" on page 201
Remote command:
TRIGger:A:LIN:TYPE on page 423 (IDDT)
TRIGger:A:LIN:DCONdition on page 425
TRIGger:A:LIN:DATA on page 425
No. of Bytes ← Data Setup ← Identifier and data
Defines the length of the data pattern - the number of bytes in the pattern.
Remote command:
TRIGger:A:LIN:DLENgth on page 426
11.6.4 LIN Label List
Label lists are protocol-specific. A LIN PTT file contains two values for each identifier:
"ID / Addr": hexadecimal identifier value
"Symbolic label": symbolic name for the identifier
Example: LIN PTT file
# ----------------------------------------------------------------------------
@FILE_VERSION = 1.0
@PROTOCOL_NAME = lin
# ----------------------------------------------------------------------------
# Labels for LIN protocol
# Column order: Identifier, Label
# ----------------------------------------------------------------------------
# Labels for standard addresses
LIN (Option R&S RTM-K3)