EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Page 471

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
456User Manual 1317.4726.02 ─ 01
STATus:QUEStionable:MASK:CONDition?.......................................................................456
STATus:QUEStionable:ENABle.......................................................................................456
STATus:QUEStionable:COVerload:ENABle......................................................................456
STATus:QUEStionable:LIMit:ENABle...............................................................................456
STATus:QUEStionable:MASK:ENABle.............................................................................456
STATus:QUEStionable[:EVENt]?.....................................................................................457
STATus:QUEStionable:COVerload[:EVENt]?....................................................................457
STATus:QUEStionable:LIMit[:EVENt]?.............................................................................457
STATus:QUEStionable:MASK[:EVENt]?...........................................................................457
STATus:QUEStionable:NTRansition................................................................................457
STATus:QUEStionable:COVerload:NTRansition...............................................................457
STATus:QUEStionable:LIMit:NTRansition........................................................................457
STATus:QUEStionable:MASK:NTRansition......................................................................457
STATus:QUEStionable:PTRansition.................................................................................457
STATus:QUEStionable:COVerload:PTRansition................................................................457
STATus:QUEStionable:LIMit:PTRansition.........................................................................457
STATus:QUEStionable:MASK:PTRansition......................................................................457
STATus:PRESet
Resets all STATUS:QUESTIONALBLE registers.
Usage:
Event
STATus:QUEStionable:CONDition?
STATus:QUEStionable:COVerload:CONDition?
STATus:QUEStionable:LIMit:CONDition?
STATus:QUEStionable:MASK:CONDition?
Returns the contents of the CONDition part of the status register to check for questionable
instrument or measurement states. Reading the CONDition registers does not delete the
contents.
Return values:
<Condition> Condition bits in decimal representation
Range: 1 to 65535
Increment: 1
Usage: Query only
STATus:QUEStionable:ENABle <Enable>
STATus:QUEStionable:COVerload:ENABle <Enable>
STATus:QUEStionable:LIMit:ENABle <Enable>
STATus:QUEStionable:MASK:ENABle <Enable>
Sets the enable mask that allows true conditions in the EVENt part to be reported in the
summary bit. If a bit is set to 1 in the enable part and its associated event bit transitions
to true, a positive transition occurs in the summary bit and is reported to the next higher
level.
Command Reference

Table of Contents