Search
R&S
®
RTM20xx
140User Manual 1317.4726.02 ─ 01
Edge
Sets the slope to be found:
●
"Rising" to search for rise time
●
"Falling" to search for fall time
●
"Both" to search for rise and fall time
Remote command:
SEARch:TRIGger:RISetime:SLOPe on page 365
Upper level
Sets the upper voltage threshold. When the signal crosses this level, the rise/fall time
measurement starts or stops depending on the selected slope.
Remote command:
SEARch:TRIGger:LEVel:RISetime:UPPer on page 365
Lower level
Sets the lower voltage threshold. When the signal crosses this level, the rise time mea-
surement starts or stops depending on the selected slope.
Remote command:
SEARch:TRIGger:LEVel:RISetime:LOWer on page 365
Comparison
Sets how the measured rise or fall time is compared with the given limit(s).
"Greater than"
Finds rise/fall times longer than the given "Rise/Fall time".
"Lower than"
Finds rise/fall times shorter than the given "Rise/Fall time".
"Equal"
Finds rise/fall times equal to the reference "Rise/Fall time" if "Varia-
tion" Δt = 0.
If "Variation" ≠ 0, the setting finds rise/fall times within the range time
± Δt.
See also: Variation.
"Not equal"
Finds rise/fall times unequal to the reference value if "Variation" Δt = 0.
If "Variation" ≠ 0, the setting finds rise/fall times outside the range time
± Δt.
Remote command:
SEARch:TRIGger:RISetime:RANGe on page 365
Rise/Fall time
Sets the reference rise or fall time, the nominal value for comparisons.
Remote command:
SEARch:TRIGger:RISetime:TIME on page 366
Reference for Search Menu