Remote Control
R&S
®
RTM20xx
361User Manual 1317.4726.02 ─ 01
SEARch:SOURce <SearchSource>
Selects the waveform to be analyzed. One of the active channel, math, and reference
waveforms can be searched.
Parameters:
<SearchSource> CH1 | CH2 | CH3 | CH4 | MA1 | MA2 | MA3 | MA4 | RE1 | RE2 |
RE3 | RE4
*RST: CH1
Example:
chapter 15.3.2.1, "Searching for a Pulse of Specified Width",
on page 460
Firmware/Software: FW 03.400
SEARch:GATE:MODE
Defines the search area. If the search is performed on a running acquisition series, the
instrument analyzes the displayed data. The search on a stopped acquisition analyzes
the contents of the memory.
Parameters:
<GateMode> OFF | DISPlay | ABSolute
OFF
Running acquisition: all waveform samples that are displayed on
the screen.
Stopped acquisition: all data samples that are stored in the mem-
ory.
DISPlay
Search is restricted to the time range of the display.
ABSolute
Search is restricted to the time range defined by SEARch:
GATE:ABSolute:START and SEARch:GATE:ABSolute:STOP
.
SEARch:GATE:ABSolute:START <StartTime>
Sets the start time of the search area in relation to the trigger point if SEARch:GATE:
MODE on page 361 is set to ABSolute.
Parameters:
<StartTime> Default unit: s
SEARch:GATE:ABSolute:STOP <StopTime>
Sets the end time of the search area in relation to the trigger point if SEARch:GATE:
MODE on page 361 is set to ABSolute.
Parameters:
<StopTime> Default unit: s
Command Reference