Remote Control
R&S
®
RTM20xx
376User Manual 1317.4726.02 ─ 01
BUS<b>:POSition <Position>
Sets the vertical position of the decoded bus signal in divisions on the sreen.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Position> Range: 4 to -4
Increment: 0.02
*RST: -3.5
Default unit: div
BUS<b>:
RESult
Displays or hides the table of decode results.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<ShowResultTable> ON | OFF
BUS<b>:LIST?
Returns the contents of the frame table in block data format.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Return values:
<DataTable> Block data
Usage: Query only
BUS<b>:LIST:SAVE <FilePath>
Saves the decoded data (frame table) to the specified CSV file (comma-separated list).
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Setting parameters:
<FilePath> String containing the storage device, path, and file name