Remote Control
R&S
®
RTM20xx
330User Manual 1317.4726.02 ─ 01
CURSor<m>:TRACking[:STATe] <State>
If set to ON, the V-Marker cursor measurement is enabled.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST: OFF
CURSor<m>:
X1Position <Xposition1>
CURSor<m>:X2Position <Xposition2>
CURSor<m>:X3Position <Xposition3>
The commands specify the x-positions of vertical cursor lines on the time axis. The third
cursor is only used for Ratio X measurement.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Parameters:
<Position> Range: Depends on horizontal settings.
CURSor<m>:Y1Position <Yposition1>
CURSor<m>:Y2Position <Yposition2>
CURSor<m>:Y3Position <Yposition3>
The commands specify the positions of horizontal cursor lines on the y-axis. The third
cursor is only used for Ratio Y measurements.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Parameters:
<Position> Range: Depends on various other settings.
CURSor<m>:YCOupling <Coupling>
CURSor<m>:XCOupling <Coupling>
If enabled, the cursors of a set are coupled so that the distance between the two remains
the same if one cursor is moved.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Parameters:
<Coupling> ON | OFF
*RST: OFF