EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Page 400

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
385User Manual 1317.4726.02 ─ 01
Return values:
<StopTime> Range: depends on sample rate, record length, and time
base
Increment: depends on the time base
Default unit: s
Usage: Query only
BUS<b>:SPI:FRAME<n>:WORD<o>:MOSI?
Returns the data value of the specified word on the MOSI line.
Use this command if only one line is defined.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
<n> *
Selects the frame (1...n)
<o> *
Selects the word number (1...m)
Return values:
<Data> Decimal value of the data word
Usage: Query only
BUS<b>:SPI:FRAME<n>:WORD<o>:MISO?
Returns the data value of the specified word on the optional MISO line.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
<n> *
Selects the frame (1...n)
<o> *
Selects the word number (1...m)
Return values:
<Data> Decimal value of the data word
Usage: Query only