Loading...
Remote Control
R&S
®
RTM20xx
394User Manual 1317.4726.02 ─ 01
BUS<b>:I2C:FRAMe<n>:STOP?
Returns the end time of the specified frame.
Suffix:
<b>
.
1..4
Selects the bus.
<n> *
Selects the frame.
Return values:
<EndTime> Range: depends on sample rate, record length, and time
base
Increment: depends on the time base
Default unit: s
Usage: Query only
BUS<b>:I2C:FRAMe<n>:AACCess?
Returns the address acknowledge bit value for the indicated frame.
Suffix:
<b>
.
1..4
Selects the bus.
<n> *
Selects the frame.
Return values:
<Acknowledge> INComplete | ACK | NACK | EITHer
Usage: Query only
BUS<b>:I2C:FRAMe<n>:ACCess?
Returns the transfer direction - read or write access from master to slave.
Suffix:
<b>
.
1..4
Selects the bus.
<n> *
Selects the frame.
Return values:
<Access> INComplete | READ | WRITE | EITHer | UNDF
INComplete
The frame is not completely contained in the acquisition.
UNDF
Access is not defined.
Usage: Query only