EasyManua.ls Logo

Rohde & Schwarz SMC100A - Page 172

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMC100A
157Operating Manual 1411.4060.32 ─ 09
The buttons are used to reset the RF sweep (all sweep modes) or to execute the RF
sweep ("Single" mode).
The sweep range, sweep spacing and dwell time are set in the bottom section.
You can configure the sweep range of the RF sweep in two ways, either by entering
the "Start" and "Stop" values or by entering the "Center" frequency and the "Span".
The two sets of parameters correlate as follows:
"Start Freq" = "Center Freq" - "Span"/2
"Stop Freq" = "Center Freq" + "Span"/2
"Center Freq" = ("Start Freq" + STOP FREQ)/2
"Span" = "Stop Freq" - "Start Freq"
RF Frequency Sweep Settings
To access the sweep dialog, select "RF > configure > Sweep/List > RF Frequency
Sweep".
In these dialogs you can configure the corresponding sweep signal.
State - Frequency Sweep
Activates RF sweep mode.
Note: Activating the RF sweep automatically deactivates other sweeps.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 294
Mode - RF Frequency Sweep
Selects the RF frequency sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMC100A

Related product manuals