EasyManua.ls Logo

Rohde & Schwarz SMC100A - Page 173

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMC100A
158Operating Manual 1411.4060.32 ─ 09
"Auto"
Generates a continuously repeating sweep signal immediately after
activating the sweep mode.
The sweep steps are performed atomatically, controlled by the dwell
time, see " Dwell Time - Frequency Sweep " on page 163.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR AUTO
SOUR:FREQ:MODE SWE
"Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
To trigger the sweep, use "Execute Single Sweep" button, or the cor-
responding remote control commands, for example *TRG.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMC100A