EasyManua.ls Logo

Rohde & Schwarz SMC100A - Page 307

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMC100A
292Operating Manual 1411.4060.32 ─ 09
f
CENTer
= (f
STOP
+ f
STARt
)/2
Note: You can select any frequency within the setting range. The range is defined with
the parameters [:SOURce<hw>]:FREQuency:STARt and [:SOURce<hw>]:
FREQuency:STOP.
A defined offset and the multiplier factor affect the sweep frequency range and there-
fore all correlated parameters. The set frequencies are only absolute values, if the off-
set = 0 and the multiplication factor = 1. The multiplier multiplies the frequencies
accordingly, and the offset ≠ 0 shifts the frequencies corresponding to the set value.
300 kHz * f
MULTiplier
+ f
OFFSet
... f
max
* f
MULTiplier
+ f
OFFSet
Parameters:
<Center> float
Range: full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST: depends on model
Example:
FREQ:CENT 400 MHz
sets the center frequency for the frequency sweep to 400 MHz.
FREQ:SPAN 200 MHz
sets a span of 200 MHz. This sets the sweep range to 300 MHz
to 500 MHz.
Manual operation: See " Center Freq - Frequency Sweep " on page 161
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal.
In CW mode, see FREQ:MODE CW|FIXed, the instrument operates at a fixed fre-
quency.
In sweep mode FREQ:MODE SWE, the value applies to the sweep frequency and the
instrument processes the frequency settings in defined sweep steps.
You can enter either a numerical frequency value, of decrease or increase the current
frequency step by step with FREQ UP and FREQ DOWN. The frequency is then
increased or decreased by the value [:SOURce<hw>]:FREQuency:STEP[:
INCRement] in FREQ:STEP:MODE USER.
Note:
A defined offset and the multiplier factor affect the sweep range and therefore all corre-
lated parameters. The set frequencies are only absolute values, if the offset = 0 and
the multiplication factor = 1. The multiplier multiplies the frequencies accordingly, and
the offset ≠ 0 shifts the frequencies corresponding to the set value.
The actual frequency at the RF output does not change, but rather the value queried
with [:SOUR]:FREQ?, according to the formula:
f
FREQ
= f
RFout
* f
MULTiplier
+ f
OFFSet
Correlation: FREQ for FREQ:MODE SWE is linked to the sweep frequency.
SOURce Subsystem