Operating Instructions Cubis
®
MCE 25
Operating Concept
Level 1 Level 2 Level 3 Description
SETUP BALANCE DISP.DIG. − Define whether all decimal places are displayed.
− Alower display accuracy enables a faster display.
− The setting option is not available on conformity-
assessed devices.
CAL./ADJ.
“Adjust”
Define the function of the [Adjust] button.
CAL.SEQ.
“Calibration
sequence”
Define the sequence of the “Calibrate and adjust”
function.
ON Z/T
Tare / zero at power
on
Activate | deactivate automatic taring when starting
the device.
ISOCAL Configure the isoCAL function.
CAL.UNIT
“Calibration weight
unit”
Define the weight unit for the calibration weight.
GEN.SERV.
“General services”
MEN.RESET
“Menu reset”
Reset the menu to factory settings.
DEVICE RS-232
“COM-RS232,
9-pin”
DAT.REC. Set the communication protocol for the connected
device.
BAUD Set the baud rate for the connected device.
PARITY Set the parity for the connected device.
HANDSHK. Set the handshake between the balance and the
device.
DATABIT Set the number of data bits.
PC-USB
“USB-B (Device/
Slave)”
DAT.REC. Set the communication protocol for the connected
PC.
BAUD Set the baud rate for the connected PC.
PARITY Set the parity for the connected PC.
HANDSHK. Set the handshake between the device and the
connected PC.
DATABIT Set the number of data bits.
USB
“USB-A (Host/
Master)”
DEV.USED
“Device in use”
Display the device connected via USB-A.