46 Operating Instructions Cubis
®
MCE
System Settings
Parameters Setting values Explanation
DISP.DIG. ALL* “Show all decimal places”: All decimal places are shown in the display. Not
available on conformity-assessed devices.
LP.ON/OFF “Reduced by 1 decimal place for load change”:
The last decimal place on the display is switched off until stability is achieved.
DIVIS. 1 “Last decimal place of the 1st division”: The last decimal place always shows the
1st division.
MINUS 1 “Last decimal place off”: The last decimal place is switched off.
CAL./ADJ. EXT.CAL. The [Adjust] button starts an external calibration with the preset calibration
weight.
E.CAL.USR. The [Adjust] button starts an external calibration with the user-defined
calibration weight value.
INT.CAL.* The [Adjust] button starts an internal calibration.
INT.LIN. The [Adjust] button linearizes the balance with the integrated weight (not for
precision balance)
CAL.SEQ. ADJUST* Calibration and adjustment is one routine.
CAL.-ADJ. Adjustment must be started or exited manually after calibration with the
[Adjust] button.
ON Z/T ON* Activates the initial taring | zeroing.
The device is tared or zeroed after it is switched on.
OFF Deactivates the initial taring | zeroing: After it is switched on, the device shows
the value before it was last switched off.
ISOCAL OFF Switches the isoCAL function off.
NOTE TO The [isoCAL] button flashes if the balance needs to be adjusted.
The isoCAL function must be manually triggered with the [Adjust] button.
ON* Activates the isoCAL function.
The device is automatically adjusted as soon as a trigger starts the isoCAL
function.
ON W/LIN.
(only for semi-
microbalances
and analytical
balances)
Activates the isoCAL and linearization function.
The device is automatically adjusted and then linearized as soon as a trigger
starts the isoCAL function.
CAL.UNIT GRAMS* Changes the calibration weight unit to grams.
KILOGR. Changes the calibration weight unit to kilograms (not for semi-microbalances
and microbalances).
CONF.UNIT Changes the calibration weight unit to milligrams (not for precision balances
and high-capacity precision balances).
* Factory setting