48 Operating Instructions Cubis
®
MCE
System Settings
Parameters Setting values Explanation
INT.ADJ. NONE The draft shield does not perform an action if an internal
calibration | adjustment starts.
CLOSE* Draft shield closes if an internal calibration | adjustment starts.
* Factory setting
Parameters in the “Ionizer” Submenu (Only for Devices with an Ionizer)
Parameters Setting values Explanation
POWER OFF Deactivates the ionizer.
SOFT Sets the intensity of the ionization process to soft.
MEDIUM* Sets the intensity of the ionization process to medium.
STRONG Sets the intensity of the ionization process to strong.
DURATION 3...15 SEC Sets the duration of the ionization process (factory setting: 8 seconds).
* Factory setting
Parameters in the “Leveling” Submenu
Parameters Setting values Explanation
LEVEL OFF The device does not display messages on the status of the level.
NOTE* The [LEVELING] display flashes if the device needs to be leveled.
ERR.MSG. The [LEVEL.ERR.] message appears in the operating display if the device
has not been leveled.
START KEY Sets the trigger for leveling to “key”: Leveling must be started manually by
pressing the [Leveling] button.
AUTO* Sets the trigger for leveling to “automatic”: The device automatically
performs a leveling function using the integrated sensor before every
internal calibration | adjustment.
* Factory setting
Parameters in the “Motion Sensor” Submenu
Parameters Setting values Explanation
CONTROL OFF Deactivates the motion sensor: No gestures are recognized.
ON* Activates the motion sensor.
NUM.GEST. 2 GESTUR.* Activates the control of functions with 2 gestures. Gestures: Left, right
4 GESTUR. Activates the control of functions with 4 gestures. Gestures: Left, right, top,
bottom
* Factory setting