Operating Instructions Cubis
®
MCE 75
Malfunctions
Warning
Message
Fault Cause Remedy Chapter,
page
ERR 02 The device cannot be
calibrated when
starting the
calibration function
due to a zero point
error.
The device was not tared or
zeroed before calibration.
− Tare or zero the device.
− Check the preload and
set if necessary.
The device is loaded. Remove the sample from
the weighing pan.
ERR 10 Taring is not possible. The device cannot be
manually tared because an
application program has the
tare memory reserved.
To release the tare memory:
Exit the application program
with the [Back] button.
ERR 11 The weight value
cannot be saved in
the tare memory.
The weight value is negative
or “zero”.
− Check the sample being
weighed.
− Zero the device before
placing the sample on
the balance if necessary.
ERR 520 to
ERR 527
The component
replacement is faulty.
The software update
could not be
completed.
The firmware versions of the
available components do
not match.
Check the device’s
firmware.
Perform a software update. 9.6, 72
If the problem occurs again:
Contact Sartorius Service.