y The procedure will be executed.
t Wait until calibration/adjustment has been completed.
y The results are displayed after the procedure is completed.
t To return to operating mode, select Back.
Calibration/Adjustment Using External
Calibration Weight
h
An external calibration weight is required for this function.
Please note the tolerance of the calibration weight being used.
t If required, switch to operating mode.
t Make sure that the weighing pan is empty.
J t Press TARE to tare the balance.
t Select Cal./Adj.
t Place the calibration weight on the balance.
t Select External cal./adj. with factory-defined weight or, if this
option is already marked, select Start.
y Calibration is carried out (approx. 15 to 20 sec.)
y After calibration, the determined deviation is displayed.
t To adjust the balance, select Adjust.
t To complete calibration without adjustment, select Exit.
Calibration/Adjustment
Cubis MSU User Manual 117

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals