System Settings (Menu)
Cubis MSU User Manual 37
Calibration/Adjustment Data
Settings are divided into the following submenus:
– Define calibration/adjustment functions
– Define external calibration weights
– Set fully autom. adjustment function isoCAL
t Configure the settings for all submenus.
t To save the settings, select Save.
h
Factory settings are marked with an asterisk (*).
Define Calibration/Adjustment Functions
Cal./adj. key: Function: Block key / *Configure key selection menu / Select
fixed 1-key function
Cal./adj. selection menu: Select which calibration/adjustment functions
are listed when the Cal./adj. key is touched.
Cal./adj. sequence: Options: *Calibrate, then adjust automatically / Calibrate,
then adjust manually
Cal./adj. unit: Options: *Gram / Kilogram / Defined via data record
Define External Calibration Weights
No. of ext. cal. weights: Enter the number.
Ext. cal. weight n: Enter the data for each calibration weight: weight value
(in grams), ID, certificate, validation
Set Fully Automatic Adjustment Function isoCAL
isoCAL: Define the function setting: Off / Show alert levels, manual start /
*On, automatic execution
isoCAL alert level: *isoCAL status field only / Alert message, repeating /
Error message, adjustment required
isoCAL temperature: Enter temperature difference (in Kelvin).
isoCAL interval time: Enter interval time in hours.