System Settings (Menu)
40 Cubis MSU User Manual
y The programmed action is displayed in the list of actions.
You can edit them again at any time here.
Device Information
You can view various information here.
t Select the information that you want to view.
Display Device Information
Here you can find all information about the manufacturer, model, serial number,
host name, and IP address used.
t To view the parameters for SQmin and DKD, select SQminDKD.
t To view the version of the balance, display and control unit, and application
software, select Version.
t To view the weighing ranges and digits (resolution), select Ranges.
Service Information/Hotline
Service contact information is displayed here along with the hotline number
and maintenance contract and interval information.
Calibration/Adjustment Log File
The log file is displayed here with the results of all calibration and adjustment
routines.
View Audit Trail
All changes to the menu, user profiles, and task profiles are logged in the audit
trail. These logs will include the following information: user, profile, parameter ID,
parameter, old and new setting.