Totalizing
Cubis MSU User Manual 99
Totalizing
Purpose: Adding up weight values and calculating values. The weight values of all individual
components are recorded and can be logged. The weight values and the component
counter are saved in protected memory so that an interrupted totalization process
can be continued after the balance has been turned off or after a power failure.
Options: You can save a maximum of 65,535 components.
User rights: All settings for the “Totalizing” application are saved under the active user profile.
Each user can make their own settings for this application. For this reason, make
sure that the desired user profile is selected before you start.
Results: When a user creates a local totalizing task, the results are only saved locally for that
user. The results of global tasks, however, are available to all users and
the administrator.
Combination options: Calculation, averaging, weighing in percent, parts counting, checkweighing, mass
unit conversion, second tare memory, individual identifiers (SQmin function, DKD
measurement uncertainty)
Configuring a Task: Totalizing
D
(* = factory setting)
t Define how the weight value will be stored.
*Manually: The weight value is stored as soon as the corresponding cursor softkey
is pressed.
Automatically: first value at stability: The weight value is applied auto-
matically as soon as the balance has stabilized after filling and the value exceeds
the minimum load.
Automatically: last value at stability: The last weight value with stability
is applied automatically, as soon as the balance is unloaded and the value falls
short of the minimum load. If stability was reached repeatedly between measure-
ments, the last stable weight value will be stored.
You must now select the minimum load to be applied:
(none, 10, 20, ...*100 .. 1000 digits).
Weighing Mode with Automatic Storage:
t Select the weighing mode to be used for the weight values.
Weighing in: Initial weighing of samples in a container.
Final readout: Final weighing of samples in a container.

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model