EasyManuals Logo

Sartorius Cubis Series User Manual

Sartorius Cubis Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
Statistics
Purpose: Saving and statistical analysis of weight values and other calculated values.
The values are generated as results:
Transaction counter
Sum of all values
Average value
Standard deviation
Lowest value (minimum)
Highest value (maximum)
Difference between maximum and minimum
Options: You can save a maximum of 65,535 components.
Combination options: Density determination, calculation, averaging, weighing in percent, parts counting,
timer-controlled functions, checkweighing, mass unit conversion, second tare
memory, individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring a Task: Statistics
D
* = Factory setting
t Define how the weight value will be applied.
*Manually: The weight value is applied as soon as the corresponding button
is touched.
Automatically: first value at stability: The weight value is applied auto-
matically as soon as the balance has stabilized after filling and the value exceeds
the minimum load.
Automatically: last value at stability: The last weight value with stability
is applied automatically, as soon as the balance is unloaded and the value falls
short of the minimum load. If stability was reached repeatedly between measure-
ments, the last stable weight value will be stored.
You must now select the minimum load to be applied:
(none, 10, 20, ...*100 ... 1000 digits).
Weighing Mode with Automatic Storage:
t Select the weighing mode to be used for the weight values. Weighing in:
Initial weighing of samples in a container.
Final readout: Final weighing of samples in a container.
78 Cubis MSU User Manual
Statistics

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals