Cubis MSU User Manual 97
Timer-controlled Functions
Timer-controlled Functions
Purpose: This application is used to trigger individual balance functions automatically
at a specific time or after specific time intervals. The trigger time or interval must
occur within 24 hours (1 day).
Options: You can select the following functions:
- Acoustic signal
- Hold display (freeze)
- Automatic printout of the display value
- Transfer of components for totalizing, formulation, or statistics
Example: In order to determine the amount of evaporation of a sample over time, you can
place the sample on the weighing pan and print out the weight value at fixed
intervals (e. g., every two minutes).
Combination options: Totalizing, statistics, formulation, mass unit conversion, second tare memory,
individual identifiers (SQmin function, DKD measurement uncertainty)
Configuring a Task: Timer-controlled Functions
D
t Define which function should be timer-controlled.
*Acoustic signal (beep tone): The balance makes a beeping sound.
Lock in readout (hold display): The measured weight value remains
on the display for the defined time period.
Automatic printout of values: The displayed value is printed
at the defined time.
t Select the mode.
*Timer set to intervals: The function is triggered at regular intervals
(e.g., every 30 sec.)
Timer set to a specific time: The function is triggered at the fixed time
(after the start of the application) (e.g., at 08:00:00).
t Specify when the time and/or interval should be entered.
Enter fixed time: You must enter this time and/or interval now.
*Time setting can be changed during process: The time and/or interval
is only entered when the task is carried out.

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals