System Settings (Menu)
36 Cubis MSU User Manual
Level the Balance (Q-Level)
It is essential for exact weighing results that the balance is absolutely level.
The front leveling feet can be used to level out small tilts in the floor. Depending
on the model, the balance is equipped with either motorized or manually operated
leveling feet.
All Cubis balances are equipped with the automatic Q-Level feature. An integrated
sensor detects the alignment of the balance and triggers a warning message when
leveling is required. You can configure this warning message and the automatic
leveling function (see System Settings > Balance Parameters).
When the isoCAL function is activated, the isoCal trigger is set after manual
leveling. Balances with motorized leveling feet are leveled automatically prior to
each new calibration/adjustment.
When “Level me!” appears on the display, you must level the balance.
t Make sure that the weighing pan is empty.
t Select the option Level the balance from the menu.
Balance with Motorized Leveling Feet
t To start the leveling function, select Next.
t Follow the instructions on the display.
Balance with Manually Operated Leveling Feet
t Follow the instructions on the display.
3
After each leveling routine, the balance must be calibrated/adjusted
(see Section “Calibration and Adjustment”).