Parts Counting
Purpose: This application is used to determine the number of objects which each have
approximately equal weight. This saves you from having to count individual parts.
Options: You can calculate or enter the piece weight of individual objects.
Combination options: Checkweighing, timer-controlled functions, totalizing, statistics, formulation,
mass unit conversion, second tare memory, individual identifiers (SQmin function,
DKD measurement uncertainty)
Configuring a Task: Parts Counting
D
t Select the method that you want to use.
*With calculation of reference piece weight: The piece weight is deter-
mined by weighing a known reference piece quantity.
With input of reference piece weight: The piece weight is entered as a
numerical value.
With calculation of the reference piece weight:
t Define how the reference piece count should be entered.
Enter fixed reference quantity: You must enter this reference piece count
nRef now.
*Reference quantity can be changed during process: The reference
piece count is entered when the task is executed.
When entering the piece count:
t Define how the piece weight should be entered.
Enter fixed piece weight: You must enter this piece weightwRef now.
*Piece weight can be changed during process: The piece weight is entered
when the task is carried out.
Parts Counting
Cubis MSU User Manual 105

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model