Installation Instructions
Adjustment to the ambient conditions:
Set up the device on a stable, even surface that is not exposed to vibrations
(e.g. weighing stone).
Ensure the device is located in a place where excessive heat cannot build up.
Maintain free access to the device at all times.
Avoid adverse influences at the place of installation:
Heat (heater or direct sunlight)
Direct drafts from open windows, air conditioning systems, and doors
Extreme vibrations during weighing
Excessive moisture
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when
it is brought into a substantially warmer area. To avoid the effects of condensation,
condition the balance for about two hours at room temperature, leaving it
unplugged from the power supply.
Follow all warnings and safety precautions from the respective section
of this manual.
Setting the Angle for the Display and Control Unit
The angle of the display and control unit can be adjusted by the user in order to
ensure optimal readability of the weight values on the display at all times.
t The display and control unit can be tilted as desired.
Power Connection
t Check the voltage rating and plug design.
Make sure that the voltage rating printed on the manufacturer’s ID label
is identical to that of your local line voltage.
If the stated supply voltage or the plug design of the power cord does
not comply with your country’s standard, please inform the nearest Sartorius
representative or your dealer.
The power connection must be made in accordance with the regulations appli-
cable in your country.
In order to connect the equipment to the power supply (protection class 1),
use a correctly installed wall outlet with a protective grounding conductor (PE)
and a fuse of maximum 16 A.
If a main feeder from the ceiling is required or if installation of a CEE plug
is necessary, these must be carried out by a qualified electrician.
Only use original Sartorius power cables.
16 Cubis MSU User Manual
Getting Started
°C

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals