Also, configure the following Jumpers on the optional RF Adaptation Board”.
RF ADAPTATION BOARD (A3514--04)
JUMPER POSITION FUNCTION
Set all jumpers +24 VDC for the inputs PREP ORDER, RAD ORDER, and FLUORO ORDER
JP1, JP3, JP4,
JP8, JP9,
J
P
,
J
P
,
Remove all jumpers
230 VAC for the inputs PREP ORDER, RAD ORDER, and FLUORO ORDER
J
P
,
J
P
,
JP13, JP14
Set only JP1,
JP8 and JP12
115 VAC for the inputs PREP ORDER, RAD ORDER, and FLUORO ORDER
J
P
Set Generator +24 VDC for PREP / RAD / FLUORO ORDER
JP2
Removed External supply for PREP / RAD / FLUORO ORDER
J
P
A ZOOM 1 output selected from Generator (--9 IN SEL)
JP5
B ZOOM 1 output selected from Table or external control
J
P
A ZOOM 2 output selected from Generator (--6 IN SEL)
JP6
B ZOOM 2 output selected from Table or external control
J
P
A ZOOM 3 output selected from Generator (--4 IN SEL)
JP7
B ZOOM 3 output selected from Table or external control
J
P
A LIH output through a N.O. contact
JP11
B LIH output through a N.C. contact
J
P
A LIH output selected from an external enable signal
JP15
B LIH output selected for Last Image Hold function
J
P
A EXP ON/END output active for only RAD exposure
JP16
B EXP ON/END output active for Fluoro and RAD exposure
J
P
A For EXP ON output active along the RAD exposure
JP17
B For EXP END output active about 50 ms pulse at the end of the RAD exposure
J
P
A For ABC Window adjustment
JP18
B For normal operation
A Pulsed Fluoro sync. activated with the negative edge of Exp. Sync+
JP19
B Pulsed Fluoro sync. from composite video signal (video in)
C Pulsed Fluoro sync. activated with the positive edge of Exp. Sync+
A For ABC OUT signal from composite video signal (video in)
JP20
B For ABC OUT signal from a negative external ABC signal
C For ABC OUT signal from a positive external ABC signal
J
P
A ABC OUT signal generated from composite video signal or external ABC signal
JP21
B ABC OUT signal coming directly from the Image System
J
P
Set ABC OUT signal generated from composite video signal or external ABC signal
JP22
Removed ABC OUT signal coming directly from the Image System
J
P
Set Normal position (composite video signal referenced to the Generator ground)
JP23
Removed To reduce noise (composite video signal isolate from Generator ground)
J
P
A Normal position (Fluoro order from the Table sent directly to the Generator)
JP24
B For Fluoro order enable (Fluoro order from the Table sent to the Image System)