HF Series Generators
Maintenance
MA-1004R4
9
7. If Fluoro is installed, select a workstation for Fluoro operation and
observe:
G Fluoro parameters are displayed on the Fluoro Display.
G Change the Fluoro kV and observe that changes are correctly
displayed.
G Check selection of another functions related to Fluoro if they are
present (ABC, PPS, etc.).
2.5.1 TOUCH SCREEN SENSOR CALIBRATION
This calibration procedure only applies to Touch Screen Console.
If required to calibrate the Sensor of the Touch Screen because the buttons can
not be properly selected or because the Compact Flash has been changed,
perform the next procedure:
CALIBRATION FOR AN “ELO” TOUCH SCREEN SENSOR
1. Enter in Service Mode and press the Software Upgrade button.
2. Press the Start--Windows button on the keyboard connected to the
Touch Screen Console, then select (double-click):
Settings / Control Panel / EloTouchscreen”.
3. Execute the Align program and follow the process touching on the
indicated places. Click on “Yes” and “OK”, then close the “Control Panel”.
4. Return to the Application through Start and select: Programs / Start up
(select the first one) / Console”.
CALIBRATION FOR A “3M” TOUCH SCREEN SENSOR
1. Enter in Service Mode and press the Software Upgrade button.
2. On the PC Desktop, press the“Start -Windows button on the keyboard
connected to the Touch Screen Console, then select: Programs / UPDD
/ Calibrate”.
3. Execute the Calibrate program and follow the process clicking on the
indicated places.
4. When finish this calibration, come back to the Application by entering
again in Start and select: Programs / Start up (select the first one) /
Console”.
Note .