HF Series Generators
Configuration
CF-1018R3
7
AEC ADAPTATION BOARD (A3263--03)
I
O
N
C
H
A
M
B
E
R
T
Y
P
E
JUMPERS POSITION
ION CH
A
MBER T
Y
PE
JP3, JP4, JP7, JP8 JP1, JP2, JP5, JP6 JP13, JP14, JP15, JP16
IC1 = IC2 = IC3 = IC4 (Default) B B B
IC1 = IC2 = IC3 B B A
IC1 = IC2 B A A
IC1 IC2 IC3 IC4 A A A
I
O
N
C
H
A
M
B
E
R
O
U
T
P
U
T
JUMPERS POSITION
ION CH
A
MBER OUTPUT
JP9 (IC1) JP10 (IC2) JP11 (IC3) JP12 (IC4)
NO-OFFSET ADJUSTMENT (Default) A A A A
OFFSET ADJUSTMENT B B B B
TEST POINT AND POTENTIOMETER
(ONLY IF JUMPER IS IN “B” POSITION)
TP1 -- R11 TP2 -- R8 TP4 -- R2 TP12 -- R5
1.4 FLUORO CONFIGURATION
Fluoro configuration depends on position of jumpers W1 and W2 in the Fluoro
CPU Board and jumper JP4 in the Console CPU Board”, as indicated below:
JUMPERS IN FLUORO CPU BOARD
(A3213-XX)
INSERTED REMOVED
W1 ABC not enable ABC enable
W2 Always inserted (installed)
JUMPERS IN ATP CONSOLE CPU BOARD (A3024-XX)
JP4 Always in “B” position -- Camera