HF Series Generators
Configuration
CF-1018R3
29
3.3 GENERATORS WITH LV-DRAC (HIGH SPEED STARTER)
3.3.1 ANODE STATOR SELECTION
For Generators with High Speed Starter, configure NOW the X-ray Tube
Family (anode stator + insert) by setting the respective dip switches 3243SW1
(pos. 4 to 8) and / or 3243SW2 (pos. 4 to 8) on the Control DRAC Board. (Refer
to “LV-DRAC - Digital Rotating Anode Controller” document ).
Configuration of these dip switches automatically determines the appropriate
Starting and Running Stator Voltage and Rotor Acceleration Time of the
selected Tube Family.
3.3.2 PROGRAMMING OF RAD FILAMENT SETTING TIME AND FLUORO FILAMENT
HOLD-OVER TIME
With High Speed operation:
-- Rotor Acceleration Time is related to the X-ray Tube Family
configured on the Control DRAC Board (LV-DRAC).
-- Fluoro and Spot Film Rotor Hold-over Times are configured on
the Control DRAC Board (LV-DRAC).
-- Rad Filament Setting Time and Fluoro Filament Hold-over Time
are configured with dip switch 3000SW2 per this instruction.
Note .