HF Series Generators
Configuration
CF-1018R3
23
SECTION 3 X-RAY TUBE SELECTION
3.1 X-RAY TUBE INSERT PROTECTION CURVES
In order to properly select the X-ray Tube Insert Protection Curves for the Tubes
connected to the Generator, perform the following procedure:
1. Enter in c alibration mode by pressing the Manual Calibration button on
the GSM program. Select one of the workstation (WS) related to the
X-ray Tube to be configured. Then press the OK button.
2. Select the respective memory location, E02 for Tube-1 or E18 for Tube-2
(memory location is shown on the mA Display).
3. Identify in Section 4 X-ray Tube Data”, the X-ray tube that is being
installed and note its tube type number.
4. Set the tube number by pressing the “Increase” or “Decrease” calibration
buttons until the correct number is showed on the Calibration Value
panel.
5. Store the value by pressing the Store button.
6. Verify that the tube code (ID) shown in the mAs Display is the same of
the tube code listed in Section 4 X-ray Tube Data”. The tube code (ID)
can be only read for the selected X-ray Tube after pressing the Store
button.
7. If required, repeat this procedure for the other X-ray Tube.
8. Exit from calibration mode.
Record configuration data for E02 and E18 in the Data Book.
Note .