HF Series Generators
Installation
IN-1052R0
30
GENERATOR SERIAL CONSOLE
For systems using a Serial Console, only a Serial Interconnection Cable
(A7066--xx) from J5 of the Generator Cabinet must be connected to the J1 of
the Serial Console. (Refer to Section 5.2 -- Maps 54301052, A6188--03).
3.4.2 COLLIMATOR ERROR SIGNAL (OPTIONAL)
This option must be requested on the customer order. For systems using an
Automatic Collimator, the X-ray exposure can be inhibited due to a Collimator
Error. This signal can be used only for one Automatic Collimator in the system,
special interconnection should be required when two C ollimators are present.
Connect two wires from the Collimator to Terminal Block 4TS3-20 (Collimator
Error signal) and 3TS1-18 (or another GND in this Terminal Block). Collimator
Error signal goes (through the Interface Cable) from 4TS3-20 in the Generator
Cabinet to terminal J2-6 in Connector J2 of the ATP Console C PU Board. If the
system is not provided with Automatic C ollimator connect 4TS3--20 to GND
(3TS1--18).
3.4.3 DOOR INTERLOCK SIGNAL
Connect two wires from the Room Door Interlock Switch(es) to Terminal Block
3TS1-22 (Door signal) and 3TS1-23 (Door Rtn - gnd). If the X-ray Room is not
provided with a Door signal, place a jumper between both connections in
Terminal Block 3TS1.
At this point, proceed to perform the complete Configuration
and the Calibration procedures except for AEC. Once
Configuration and Calibration tasks have been performed,
proceed with the rest of the Installation and Calibration tasks.