SECTION 2 EXTENDED MEMORY SETTING
2.1 EXTENDED MEMORY LOCATIONS
Miscellaneous configuration and calibration data are stored in the Extended
Memory Locations. It is recommended to note the values factory stored in each
Memory Location. (RefertoTable2-1)
For generators with only one Radiographic X-ray Tube, this tube
have to be configured, calibrated and used as Tube-1.
For generators with a Fluoroscopic X-ray Tube or DSI (Digital
RAD), this tube have to be c onfigured, calibrated and used as
Tube-2.
For these generators, the value of the E17 Memory Location is not
readable as “Tube-2 -- Filament stand-by (Autocalibrated)”.
Generators with a s ingle X-ray Tube for Fluoroscopy / Spot Film
/ DSI (Digital RAD) require to store the respective values of the
Memory Locations:
-- E17: Tube-2 -- Fluoro filament setting.
-- E18: Tube-2 -- Fluoro tube type.
-- E29 and E31: Tube-2 -- Exposure Time adjustments.
-- Other required Memory Locations.